Error code E32 americomp xray machine
March 29, 2023 11:32

E32 is a filament error. Could be a blown fuse or an open filament in the tube. ...Click here to read more

GE Diamond mammography
February 10, 2022 10:17

Err 0077 Internal communication error. Collimator board not answering. Err0015 Collimator plate movement backup time reached, message sent to supervisor board/cpu board. Err0033 Program checksum error. Battery Ram. Right from the ...Click here to read more

Fill oil to Xray tube
November 04, 2019 10:07

The x-ray tube needs to be vacuum sealed to remove all air. Air bubbles can cause arcing in the tube....Click here to read more