Saeed Hashemi - NASS MedImage - DOTmed Profil

Saeed Hashemi - Profile
Ultrasound & Laser Sales & Service Manager
NASS MedImage
Ontario, Canada
DOTmed User since January 2005
Email Saeed

(51) (12)

Rétroaction globale reçue

Rétroaction globale donnée

Overall Rating (5 ratings)

Details

See All Feedback Given

Postes récents de forum

VBEAM PERFECTA
November 15, 2019 09:29

Hi Cameron, Low energy can be due to lots and lots of parts and reasons. Laser Head, Reflector Mirrors, Lamp, DYE Kit, Head energy monitors, HVPS, Delivery system, .... The answer to your question will be too complicated and lon...Click here to read more

VBEAM PERFECTA
November 13, 2019 01:32

Hello Cameron, I guess you have not changed the DYE Kit for a while. DYE Kit must be changed each 12 to 18 months or 50,000 shots which comes first. The problem of your machine is Low Energy which is related to old DYE Kit. The...Click here to read more

Cutera Xeo Cool Glide Laser Error Code
August 20, 2015 04:22

E249/367 is related to broken lamp and the cost of service if you were in Toronto, Ontario, Canada would be $1100....Click here to read more

All forum posts by Saeed Hashemi