We provide sales, service, rental, and repair for ultrasound systems, probes, and C-arms. We have the largest inventory in the world. Contact us about our special offers on certified refurbished equipment! Call: 317-759-9210 Email: sales@probomedical.com.
North America's trusted source for ultrasound probe repair and certified refurbished OEM probes. Free evaluations, 1-year warranty, fast turnaround for GE, Philips, Sonosite, Canon, Toshiba, and more. Call (224) 265-7640 / Sales@Lenard.Healthcare
Faulty Phenomenon Malfunction: during operation, errors are observed on all sensors throughout the entire depth of scanning. There is no possibility of full operation. Troubleshooting and Solutions... voir plus
Faulty Phenomenon Malfunction: When attempting to start a full-depth scanning investigation, interference occurs in the form of vertical stripes. In addition, the error "Error on Power Management"... voir plus
Faulty Phenomenon Malfunction: Upon loading the software support, an error "NATHAN LUT error" appeared. Troubleshooting and Solutions Malfunction: Upon loading the software, an error "NATHAN LUT... voir plus
Faulty Phenomenon Malfunction: during operation of the device in Doppler mode in the area of interest, colored noise appears. Troubleshooting and Solutions Malfunction: during operation of the... voir plus
Faulty Phenomenon Malfunction: during operation, errors occur on any sensors at various depths of scanning. Work is complicated. Troubleshooting and Solutions Malfunction: during operation, errors... voir plus
Faulty Phenomenon Malfunction: The cursor on the screen system did not respond to movements of the trackball horizontally (left and right), which complicated the control of the device. ... voir plus
Faulty Phenomenon Malfunction: the issue has a progressive nature, that is, earlier during the scanning process, an error occurred sporadically with the frequency of "Error: Aurora communication... voir plus
Faulty Phenomenon The procedure for updating the software version and hardware on Voluson E8 (versions BT10 to BT13) has been carried out. To install the software version BT13, a replacement BackEnd... voir plus
Faulty Phenomenon The procedure for updating the software version and hardware on Voluson E8 (versions BT10 to BT13) has been carried out. To install the software version BT13, a replacement BackEnd... voir plus
Faulty Phenomenon Malfunction: when connecting any sensor to any connector of the sensor (error PSS) solution: repair of the sensor connection plate RTF was carried out. Troubleshooting and... voir plus
Faulty Phenomenon Beamformer consists of RFM201, RFM221 (PN: KTZ303915, KTZ303916); RSX10, RSX20 (KTZ303250, KTZ303054). Error: Aurora communication failed; when turning on the device, the image... voir plus
Faulty Phenomenon Malfunction: Memory access violation, device requires restart. Repair board RFM423G. P19 (PN: KTZ304074, KTZ304142): restoration of factory firmware in EEPROM and Flash Memory for... voir plus
Faulty Phenomenon Malfunction: artifacts on the display, improper operation of ATsP and TX Pulser (hdl6v5541) in the Beamformer BF192 Assy (PN: 5357234-2). Corresponding microchips have been... voir plus
Faulty Phenomenon Malfunction: An artifact appeared in the form of a stripe on all sensors. After performing the service test BF, we detected a malfunction in the RSR board (KTZ196310, Receiver RX... voir plus
Faulty Phenomenon Malfunction: Artifacts appeared in the form of white stripes during operation with any sensor. During the diagnostic process, we detected a malfunction in the BF192 board (PN:... voir plus